Call us at 732 292-1994 or 800 867-6690 | Email us at triadscientific@gmail.com | Visite us on
PLANET BEST
Lab Instrumentation
Lab Equipment
Lab Furniture
|
(pdf download)
|
Formula
|
Description
|
GP Plus!
|
The most widely used filter in the range, primarily for solvent, organic, and alcohol removal.
|
ACI Plus!
|
Neutralizes volatile inorganic acid vapors.
|
ACR
|
Iodine and methyl iodide vapors. It is frequently used for iodination reactions with low level radioactive iodine.
|
ACM
|
Mercury vapor.
|
AMM
|
Removes vapors from dilute ammonia solutions and to remove low molecular weight amines.
|
SUL
|
Designed to remove hydrogen sulphide and low molecular weight mercaptans.
|
CYN
|
Removal of hydrogen cyanide. Many cyanide compounds will evolve HCN gas if acidified, so this fi lter is normally specified if working with any cyanide compound.
|
FOR
|
Designed to oxidize formaldehyde and glutaraldehyde fumes. It is widely used in hospital pathology laboratories.
|
ETH
|
Diethyl ether is adsorbed on activated carbon, but because of its low boiling point, local heat adsorption can reduce the capacity of the filter. Special impregnation allows a chemical reaction which increases the filter capacity.
|
EDU
|
Designed to handle chemicals normally used is a university level chemistry curriculum.
|
MIL
|
As the name implies, this filter is designed for military applications involving war gasses.
|
HEPA / ULPA
|
Powders and particulates.
|
Enhanced Filtration Technology
|
Enhanced Filtration Technology
|
Model
|
Dimensions
|
Weight (lbs/Kg)
|
||
|
External
(W x D x H)
|
Shipping
(W x D x H)
|
Net
|
Ship
|
AP60H
|
21.25" x 19.75" x 12"
540 x 502 x 305 mm |
22" x 24" x 14"
610 x 559 x 356 mm |
29 / 13
|
31 / 14
|
Air Science Model |
FUME BOX AP60H
|
Filter Weight
|
7 lbs / 3.5 kg
|
Prefilter Weight
|
1 lb / 0.4 kg
|
Airflow
|
135.9 cfm |
Face Velocity FPM
|
100 fpm
|
Construction
|
White epoxy steel filtration unit with blue enclosure, clear sides and top panels.
|
Blower
|
Low-noise ebmpapbst centrifugal fan
|
Electrical
|
120v 60hz; 220 50hz available, specify when ordering.
|
Electrical Switches
|
Main On/Off
|
FUME BOX Model
|
|
FUME BOX AP60H
|
Vertical Mount*
|
Unit mounts vertical to create an upflow airflow pattern. |
AP60V |
Source Capture Hose*
|
Converts the Fume Box to a fume extraction unit with a 6" diameter (15 cm) hose to capture pollutants at their source. Hose can attach to any vented enclosure, exhaust port on your equipment, or desk mount scoop.
|
AP60EX
|
Polypropylene Construction*
|
Units available in all polypropylene construction
|
AP60H-PP
AP60V-PP AP60EX-PP |
Spillage Tray
|
Polypropylene spillage tray, slides out for easy cleaning
|
TRAY-AP60
|