PARKER 20177 Parker Model FID-1000 Parker FID-1000 Parker Balston FID-1000 and RESTEK 20177 Gas Station used
Single unit produces UHP zero air from house compressed air and 99.9995% pure hydrogen from deionized water.
Ideal for supplying up to 5-6 FIDs.
Eliminates inconvenient and dangerous gas cylinders.
Silent operation, minimal operator attention required.
Parker gas stations provide both UHP grade hydrogen gas and zero grade air for flame ionization detectors. The system is specifically designed to supply gas to FIDs and to support flame thermionic and flame photometric detectors. The units produce zero air by purifying compressed air to a total hydrocarbon concentration of 0.1 ppm or less (measured as methane).
The hydrogen generators produce hydrogen gas from deionized water, using the principle of electrolytic dissociation of water and hydrogen proton conduction through a proton exchange membrane cell.
Specifications
Hydrogen Purity: 99.9995%
Zero Air Purity: < 0.1 ppm total hydrocarbons as methane
Max. Hydrogen Flow Rate: FID-1000: 90 cc/min, FID-2500: 250 cc/min
Max. Zero Air Flow Rate: FID-1000: 1000 cc/min, FID-2500: 2500 cc/min
Power: 120 VAC/amp, 60 Hz, 480 watts
Hydrogen Outlet Pressure: 60 psig (414 kPa)
Zero Air Outlet Pressure: 40-125 psig* (276-862 kPa)
Inlet Connection: 1/4" NPT (female)
Outlet: 1/8" compression
Dimensions: 16.5" h x 10.5" w x 17" d (42 cm x 27 cm x 43 cm)
Weight: 53 lb (24 kg)