Shimadzu UV-1700 UV Pharmaspec Spectrophotometer Shimadzu Pharmaspec Shimadzu UV1700 Pharmaspec Shimadzu UV-VIS Spectrophotometer Shimadzu 1 nm bandpass
These used systems ship checked out and ready to go- limited supply- call now!
FILE PHOTO of similar
Specifications
Spectra l bandwidth
1 nm (190 to 900 nm)
Dimensions
550W x 470D x 200 H (mm) (max. height 380 mm)
Weight
17 kg
Wavelength range
190.0 to 1100.0 nm
Wavelength display
0.1 nm increments
Wavelength setting
0.1 nm increments (1 nm increments when setting scan range)
Wavelength accuracy
±0.3 nm
Wavelength repeatability
±0.1 nm.
Wavelength Scan Rates
Wavelength slew speed: Approx. 6000 nm/min
Wavelength scanning speed:
Approx. 3000 nm/min to 10 nm/min
Lamp interchange wavelength
295.0 nm to 364.0 nm (340.8 nm)
Stray light
0.04% max. (220.0 nm NaI, 340.0 nm NaNO2)
Photometric system
Double-beam Optics
Photometric range
Absorbance: -0.5 to +3.0 Abs
Transmittance: 0.0 to 300%
Recording range
Absorbance: -3.99 to +3.99 Abs
Transmittance: -399 to +399%
Photometric accuracy
±0.004 Abs (at 1.0 Abs), using NIST930D filter
±0.002 Abs (at 0.5 Abs)
Photometric repeatability
±0.002 Abs (at 1.0 Abs)
±0.001 Abs (at 0.5 Abs)
Baseline stability
0.001 Abs/h max (700 nm, After 1hour from power on)
Baseline flatness
±0.001 Abs (1100 mm to 200 mm) After 1hour from power on
Noise level
0.002 Abs (P-P) 0.0002 A (RMS) (700nm)
Baseline correction
Automatic with computer memory, in two stages of coarse and fine
Light source
20 W halogen lamp, deuterium lamp Built-in light source automatic position adjustment
Monochromator
Aberration-corrected blazed holographic grating
Detector
Silicone photodiode
Sample compartment
Internal dimensions: 110.0W x 230.0D x 105.0H (mm)
Distance between light beams:
100.00 mm
Installation:
Fixed with two screws
Beam size:
9 x 0.5 mm
Power requirements
100, 120, 220, 240 VAC
50/60 Hz, 130 VA
Ambient requirements
Temperature: 15 to 35°C, humidity: 35 to 80%
Perkin Elmer UV/VIS/NIR Lambda 19 upgraded Perkin Elmer Lambda 19 UV VIS NIR Perkin Elmer options such as Integrating Sphere may be available, this system ships with software, computer, printer and ready to go- This double beam double monochrometer ratio recording UV Visible NEAR INFRARED Spectrophotometer, Total Useable Wavelength Range 185-3200nm, a real nice system that ships ready to go